-148
Character data
n
ot allowed
D
escrip t i o n :
-148
C haracter data
n
ot all ow ed
-200
E
xecution error
HP 8110 Α Service Guide
HP-ΙΒ
m
essage = "No MIN/MAX allowed while
w arnings are act ive"
Local message = "No MIN/MAX while warnings
are active"
An attempt was
m
ade to set α parameter to
MIN/MAX resp. t o as k for the MIN/MAX of a
p aram eter wh ile w arnin gs w ere act iv e .
This does
n
ot work becau se ot her
p
arameters on
whi ch t he MIN/MAX result depends might already
b e outside no-warn range.
HP- ΙΒ
m
essage = "No MIN/MAX allo w ed while
c
h
ec
k
s are o
ff"
Local
m
essage = "No MIN/MAX while checks are
off"
D
escrip tion :
An attempt was made to set α para m eter to
MIN/MAX resp . t o ask for the MIN/MAX of a
p
arameter while ch ec k s were s wi tched off.
Thi s does not
w
ork because other
p
aram eters on
whi ch th e MIN/MAX result depen ds mi g h t a l ready
be outside no-warn range.
HP- ΙΒ
m
essage = "No signal at CLK- IN coul d be
detected
f
or frequenc y measureme nt" .
L
ocal message = "No signal detected at
CLK-INput"
D
escript ion:
An attemp t was made to
m
eas ure the extern al
c
lock f re qu ency but no si gnal was detectable on t he
clock-in connector .
Troubles
h
ooti
n
g 4-25