-314
SA
V
e/recall memory l ost
HP-ΙΒ message = "[re m/1 οε] Setting [no]
P
attern
corrupt"
Local message = "[rem/1οε] Setting (no] Pattern
corrupt"
D
escrip ti on :
[rem/1οε] = re mote or
l
ocal, [no) = numb er input]
Save/Recall memory l ost . Indi cates that the
n onvol atile data saved by t he *SAV comm and h as
b
een lost . In this case only
t
he pattern block in
t
he
selected setting
i
s corrupt .
-314
SAVe/recall memory lost
HP 8110 Α Se
r
vice G
u
i
d
e
HP- ΙΒ
m
essage = "[rem/1οε] Setting [no] Firmware
revi s ion nr . incompatibl e"
L
ocal message = "[rem/loc] Setting [no] Firmware
rev. incompat ."
D
escrip t ion:
[rem /1 οε] = rem ote or
l
ocal, [no] = numb er input]
E
ach firmware iss ue has α revision number. The
instr um e nt is trying to read α setting with an
incomp atible revis i on numb er.
-314
S AVe/reca ll memory l ost
HP- ΙΒ
m
essage = "[re m/1 οε] Setting [no]
E
rror
check confli ct"
L
ocal
m
essage = "[rem/1οε] Setting [no]
E
rror
check confli ct"
D
escript i on:
[rem/loc] = remote or
l
ocal , [no] = numb er input]
The error check switc h in t h e sp ecified sett ing
i s OFF but the inte rnal error check is ON . The
instrument cannot recall this sett ing.
Tr
o
u
bles
h
ooti
n
g
4-4 5