150 mhz pulse generator
Hide thumbs Also See for 8110A:
Table of Contents

Advertisement

HP 811 OA Service Guide
4-46 T
r
ou
b
les
h
ooti
n
g
-314
SA
V
e/reca ll mem ory lost
HP- ΙΒ
m
essage = "[re m /Ιοε ] Settin g [no]
In strument config ur atio n incompatibl e"
L ocal message = "[rem/1 οε] Setting [no] Co n fig .
in comp at ible"
D
escript i on:
[re m/Ιοε ] = rem ote or
l
ocal, [no] = numb er input]
Ev ery stored sett ing co ntains the in str ument
co nfiguration . Th e stored config u ratio n does not
cor res pond to the act u al in strument co nfigurat ion .
-330
Se lftest failed
HP-ΙΒ
m
essage = "Se lftest error : Mi cro p rocessor
board failed"
Local m essage = "Selftest error : Micro p roc .
b
oard
failed"
D
escr ip tion :
H ardware on the Mi crop rocessorboard does not work.
-330
Se lftest failed
HP- ΙΒ
m
essage = "RO M test failed"
L
ocal m essage = "RO M test
f
ailed"
Description :
Flash EPROM test
f
ailed on Microp rocessorboard .
-330
Self test
f
ailed
HP- ΙΒ
m
essage = "RAM test fa iled"
L
ocal m essage = "RAM test
f
ailed"
D
escr iption:
Stat i c RAM test fail ed on Mi cro p rocessor b oard.
-330
Selftest failed
HP-ΙΒ
m
essage = "Crystal Refere n ce for υΡ
l
ost"
L
ocal m essage = "Cr ystal R efere nce
f
or υΡ
l
ost"
D
escription:
Α
l
oss o f cry stal refere nce has
b
een detected an d the
V CO is runnin g at approxim ately half of
m
aximum
speed, determined
f
rom an internal voltage re ference.
N
ormally the ex ter n al cry stal
f
requency is VCO
reference.