Fast transitions
display indicates, the transition times are within the specified
limits.
are given here and on the Test Record.
Note
* Under programming to 5.5 ns is permissible to meet this
specification.
Slow transitions
Set up the HP 8112A as follows:
Input Mode
Control Mode
P E R
DEL
D T Y
Transition
HIL
For the HP 8112A to the settings in the table below, measure the
displayed leading edges (LEE) and trailing edges (TRE).
Verify that the transition times do not exceed the specified limits.
Record your results on a copy of the Test Record, specified limits
are given here and on the Test Record.
Note
The oscilloscope response is very slow for measurements with PER in
the ms range. The HP 54503A or the HP 54100D oscilloscopes will
provide faster response.
NORM
Off
65 ns
50%
Linear
0.00 v
Testing Performance 8-15