Amplitude Tests; Shaper Output State Test; Burst Tests - HP 8112A Operating, Programming And Servicing Manual

50 mhz programmable pulse generator
Table of Contents

Advertisement

Amplitude Tests

Shaper Output State
Test

Burst Tests

E36
When software triggered (low), fall time of 1 ms (Range 6) out of
limits. Possible faults in trailing edge slope control circuit U13, U14
or U302. See Chapter 10.5.
E37
When software triggered (high), rise time of 9.99 ms (Range 6) out of
limits. Possible faults as in E35.
When software triggered (low), fall time of 9.99 ms (Range 6) out of
limits. Possible-faults as in E35.
E39
When software triggered (high), rise time of 10 ms (Range 7) out of
limits. Possible faults as in E35.
E41
Amplitude range 2 (9.99 V) not possible. Possible faults:
Pre-attenuator circuit
n
Amplitude control circuit U23 to U25
n
Offset control circuit U20 to U22
n
See Chapters 10.5 and 10.4.
E42
Amplitude range 1 (1 V) not possible. Possible faults as in E41.
HP 8112A set to produce a burst of 10 pulses of period 250 p.s.
E51
Burst counter supplies less than 10 pulses. Possible fault in Burst
flip-flop U201 or Burst control circuit U102 to U112. Refer to
Chapter 10.6.
Burst counter produces more than 10 pulses. Possible faults as in
E51.