9-18 Adjustment Pwcedures
HIL C
40. Set up the HP 8112A as follows:
Trigger Mode
Control Mode
P E R
DEL
WID
Transition
HIL
COMPL
DISABLE
LIMIT
41. Connect the HP 8112A output via an exact 50 R 0.1%
feedthrough, to the oscilloscope (HP 54503A).
42. Adjust A2R50 for 0 V amplitude (no signal).
43. Set the HP 8116A Signal Generator as follows:
Trigger Mode
FREQ
AMPL
DISABLE
44. Connect the signal to the HP 8112A CTRL INPUT.
45. Adjust AlR506 for minimum change in the waveform baseline.
NORM
HILC
1 ms
65 ns
500 ps
Fixed
0.0 v
Off
Off
Off
NORM
2 Hz
8 V, symetrical
Off