HP 8112A Operating, Programming And Servicing Manual page 161

50 mhz programmable pulse generator
Table of Contents

Advertisement

18.
19.
20.
Note
I
Transition times can be increased by increasing the overshoot. If
Overshoot/Transition times adjustment cannot be achieved within
specification, change values of AlC532/C409
repeat steps 1 to 15.
Take full screen display on the scope. Check that the transition
times <4.8 ns in both normal and complement modes.
Set the HP 8112A to Linear transition.
Measure the transition times are <6.2 ns.
(See table 9-l) and
Adjustment Procedures 9-9