HP 8112A Operating, Programming And Servicing Manual page 143

50 mhz programmable pulse generator
Table of Contents

Advertisement

3. Verify that the result is:
1.0 MS,DTY 50 %,HIL 1.0 V,LOL 0.0 V,MI,CTO,TI,LO,CO,BUR
DEL 65 NS,WID 100 US,DBL 200 US,LEE IO NS,TRE IO NS
4. Use the following program statements to change some instrument
5. Verify that the settings are the same as before, except for the
6. Using the oscilloscope confirm that the HP
settings and then re-read the current settings:
OUTPUT 712;"PER IO MS,
DTY IO %,
DO"
OUTPUT 712 ; "CST"
ENTER 712;B$
PRINT B$
following:
w 3
DO
PER 10.0 MS
DTY IO %
HIL 1.50 v
following form:
Square
Pulses
Period
10 ms
Duty Cycle
10%
High-Level
t1.5 v
0.0 v
Low-Level
Allocate controller memory to
receive second status string
Change settings
Request current settings
HP 8112A
Read the HP 8112A settings
Display the HP 8112A settings
8112A
output has the
Testing Performance 8-31
0001 #,