HP 8112A Operating, Programming And Servicing Manual page 129

50 mhz programmable pulse generator
Table of Contents

Advertisement

5. Record your results on a copy of the Test Record, specified limits
are given here and on the Test Record.
Note
The oscilloscope trace flatness error may affect the measurement of
pre- and overshoot.
6. Change HP 8112A settings as follows:
Transition
LEE
T R E
7. Examine the HP 8112A output on the scope display in order to
verify that the pulse characteristics do not exceed the specified
limits as identified in Figure 8-11.
8. Record your results on a copy of the Test Record, specified limits
are given here and on the Test Record.
Note
The oscilloscope trace flatness error may affect the measurement of
pre- and overshoot.
Characteristic
Preshoot
Overshoot/Ringing <&lo% of amplitude AlO mV
Settling time
Linear
6.5 ns
6.5 ns
Characteristic
Preshoot
Overshoot/Ringing 3~5% of amplitude f10 mV
Settling time
Specification
I
Testing Performance 8-17