HP 5890 Series II Plus Reference Manual page 10

Hewlett-packard reference manual
Table of Contents

Advertisement

Columns and Fittings
The HP 5890 SERIES II (hereafter referred to as HP 5890) provides
flexibility in choices among inlets, columns, and detectors through use of
liners and adapters, allowing any standard column to be used without
sacrificing performance. Additional flexibility is gained through positions
of inlets and detectors relative to each other and through the large
internal volume of the oven.
This section provides information for the following:
The column oven.
Fittings.
Liners and inserts.
ECD and TCD capillary makeup gas adapters.
The first three items must be considered before a column may be
installed properly at either an inlet or a detector. In addition, for an FID
or NPD, and depending upon the column to be installed (packed versus
capillary), the correct jet must be installed before installing the column.
Jet installation is described in Chapter 8, Preventive Maintenance.
For specific information on ordering fittings, liners, and inserts, see
Hewlett•Packard' s analytical supplies catalog.