Split/Splitless Or Split-Only Capillary Inlet Inserts - HP 5890 Series II Plus Reference Manual

Hewlett-packard reference manual
Table of Contents

Advertisement

Columns and Fittings
Inlet inserts
Note: For the liner and insert for an HP Series 530
the column is already installed, a new insert may not seat properly in the
liner; the column may prevent it from dropping completely into the liner.
If the insert does not drop completely into the liner, do not force it
(either the liner or the column may shatter); instead, remove the column,
seat the insert, and then replace the column.
5. Replace the septum and septum retainer nut.

Split/splitless or split-only capillary inlet inserts

Figure 1-10
Split/Splitless and Split-Only Capillary Inlet and Inserts
A specific inlet insert is required, depending upon the particular
sampling mode. Specific sampling modes include:
Split, for major•componentanalyses
Purged splitless, for trace•componentanalyses
Viton O-ring (Preferred)
Viton O-ring
Split Use Splitless Use
¿
capillary column, if
27