Detector Liners/Adapters - HP 5890 Series II Plus Reference Manual

Hewlett-packard reference manual
Table of Contents

Advertisement

Columns and Fittings
Inlet/detector liners/adapters

Detector liners/adapters

Figure 1-6
Liner/Adapter
Typical Installed Detector Liner/Adapter
Detectors require a liner/adapter to be installed when used with packed
metal columns (either 1/8•or 1/4•inch),and with any type of capillary
column. Normally, no liner is required with 1/4•inchpacked glass
columns, since the leg of the column itself serves as the liner.
For the FID or NPD, the correct detector jet must be installed prior to
installation of the liner. (If jets must be exchanged, see Chapter 8,
Preventive Maintenance.)