Entering Specific Zero And Gain Values - HP 5890 Series II Plus Reference Manual

Hewlett-packard reference manual
Table of Contents

Advertisement

Electronic Flow Sensing
Electronic flow sensor (EFS) calibration

Entering specific ZERO and GAIN values

Calibration values for zero and gain should be recorded when a
particular channel is calibrated. They can then be reentered through the
keyboard if necessary, without repeating the entire calibration procedure.
To enter specific zero and gain calibration values:
1. Select CALIB AND TEST mode, function 2:
GAIN A (or GAIN B) is displayed, followed by two values (the observed
flow rate through the channel and the current gain calibration value
for the channel).
Note that Channel A is assumed by default. If Channel B is to be
calibrated instead, press
2. Enter the desired gain calibration value, preceded by
than a measured flow rate.
3. Press
current zero calibration value for EFS Channel A).
Note that Channel A is assumed by default. If Channel B is to be
calibrated instead, press
4. Enter the desired zero calibration value:
Note that alternately pressing
calibration value or the gain calibration value for the given channel (A
or B).
CLEAR
Numeric Value
-
is necessary to signify entry of a gain calibration value, rather
-
:
FLOW A ZERO is displayed, followed by a value (the
ZERO
Numeric Value
.
2
ENTER
.
B
ENTER
.
B
ENTER
or
ZERO
FLOW
:
-
displays either the zero
65