HP 5890 Series II Plus Reference Manual page 235

Hewlett-packard reference manual
Table of Contents

Advertisement

Test Sample Chromatograms
Test sample chromatograms
Figure 10-11.
Detector Type
Temp 300 DEGREES C
Inlet Type Ded On-Col
Oven Track
Temp N/A
Operating Mode
Purge Time On
Purge Time Off
Oven Temp Programmed (1 ramp)
Init Temp
Init Time
Ramp
Rate
Fin Temp
Fin Time
Flow Param (EPP)
Constant Flow
Range
0
COLUMN:
Part No.
19095S(#100)
Dimensions
Sta Phase
Methyl Silicone
TCD-On-Column Capillary Inlet
HP 5890 Test Sample Operating
Conditions
TCDw/MUG
On
C
N/A
N/A
min
N/A
min
60
DEGREES C
0.5
min
20
180
1
Off
¿
530
MID; 5 M
4.23
4.82
5.38
FLOW RATES
Carrier (He)
20 +/• 1
Hydrogen
Air
Makeup (He)
5+/•1
Split Vent
Septum Purge
5+/•1
Reference (He) 37+/•1
SAMPLE:
Type
FID Sample
¿
Inj Volume
3
Part No. 18710•60170
Composition 0.03%(V/V)each
C14,C15,C16 normal paraffin
hydrocarbons in hexane.
C14
C15
C16
ml/min
N/A
ml/min
N/A
ml/min
ml/min
N/A
ml/min
ml/min
ml/min
l
235