Performance Considerations - HP 5890 Series II Plus Reference Manual

Hewlett-packard reference manual
Table of Contents

Advertisement

Detector Systems
Nitrogen-phosphorus detector (NPD)

Performance considerations

Contamination
Very little contamination can create serious NPD problems. Common
sources include:
Columns and/or glass wool treated with H PO (phosphoric acid)
Phosphate•containingdetergents
Cyano•substitutedsilicone columns (XE•60,OV•225,etc.)
Other nitrogen•containingliquid phases
Any liquid phase deactivated for analysis of basic compounds
Fingerprints
Leak•detectionfluids
Laboratory air
Contamination in the NPD can be either a positive type or a negative
type of contamination. A positive type of contamination is one that gives
a more positive offset than would normally result with a clean system.
When this happens there is a tendency to try to operate the detector with
too little power to the active element, with the final result that the
detector does not appear as sensitive as desirable (because the
temperature of the active element is actually less than normal).
If the contamination is a negative type of contamination, then there is a
tendency to quench the reactions that give rise to the specific response
desired. This quenching can be only partial, resulting in less sensitivity
than expected, or at very high contamination levels, a complete
quenching of all signals from the detector. This latter type of quenching is
(usually) only seen on sample peaks where the apex of the peak actually
goes back down toward the baseline.
132

Advertisement

Table of Contents
loading

This manual is also suitable for:

5890 series ii5890 ii series5890 ii plus series

Table of Contents