Repacking A Split Insert - HP 5890 Series II Plus Reference Manual

Hewlett-packard reference manual
Table of Contents

Advertisement

Preventive Maintenance
Liner and/or insert care

Repacking a split insert

Since, for a split insert, its packing material is discarded in cleaning, the
insert must be repacked.
Note: Repacking with small•diameterglass beads is not recommended:
they are usually contaminated with metal filings due to sieving
procedures used. If they must be used, thorough cleaning (chemical and
physical) is required.
Use a fresh, small amount of a conventional coated packing such as
2% OV•1on 100/120 mesh, Chromosorb W•HP .
Packing is held in place between plugs of silanized glass wool (Part
No. 8500•1572).
Figure 8-8
Glass
Wool
Packing
Packing Requirements, Split Insert
To ensure compatibility with standard 2•inchsyringe needles, there
should be at least 35 mm between the top of the insert and the top of
the packed section.
170
35 mm