HP 5890 Series II Plus Reference Manual page 113

Hewlett-packard reference manual
Table of Contents

Advertisement

Inlet Systems
Split/splitless capillary inlet
The split ratio is an indicator of the fraction of total sample entering the
column: the higher the value, the less sample enters the column.
For setting flow for split sampling, see Chapter 4 of the HP 5890
Operating Manual.
Verifying inlet purge status
Verify that inlet purge flow is currently on, and will remain on
throughout runs to be made in split sampling mode.
To display current inlet purge status:
If OFF is displayed, press
To display elapsed time during a run when inlet purging will be
halted:
To display elapsed time during a run when inlet purging will be
restored:
(or
For split sampling, both times should be set to 0.00 and inlet purging
must currently be ON. If necessary, this is done simply by pressing
ENTER
A
PURGE/VALVE
A
PURGE/VALVE
A
PURGE/VALVE
alone if PURGE TIME is already displayed)
ON
when PURGE TIME OFF (or ON) is displayed.
( or
)
B
to restore inlet purging.
ON
( or
)
B
TIME
( or
)
B
TIME
OFF
ON
0
113

Advertisement

Table of Contents
loading

This manual is also suitable for:

5890 series ii5890 ii series5890 ii plus series

Table of Contents