ABB Endura AZ30 series Programming Manual page 107

Integral probe and remote transmitter combustion oxygen monitor
Hide thumbs Also See for Endura AZ30 series:
Table of Contents

Advertisement

Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
Threaded entries
Gland entries (certified):
Power and signals gland entries:
Remote terminal housing cable gland entry:
Pneumatic entries (not certified):
4 fittings supplied with AutoCal options or 3 fittings and 1 blanking plug supplied with non-AutoCal options. Size
1
options:
/
in BSP for 6 mm OD pipe (with M20 cable gland option) or
4
Automatic Calibration
AutoCal hardware
Isolated solenoid valve control as standard, 24 V @ 2 W per valve*
Dedicated isolated digital inputs to monitor pressure switch contacts as standard:
voltage-free, normally closed with gas present
*For driving internal automatic calibration (AutoCal) probes or can be used to drive external automatic calibration units on
remote transmitters only.
Display and Switches
Display type
Graphical 128 x 64 pixel LCD
Display backlight
Green LED
Operator switches
4 capacitive switches (operated through the front glass)
Relay Outputs
Number
2 standard
Type
Normally closed 5 A @ 230 V AC or 30 V DC (non-inductive)
Functions
User-configurable – can be activated by one or more of the following signals:
Process alarm 1, 2, 3, 4
Calibration in progress
Calibration failed
Out of test gas 1, 2
Test gas 1 valve control
Test gas 2 valve control
Failure diagnostic
Out-of-specification diagnostic
Maintenance required diagnostic
Function check diagnostic
COI/AZ30E–EN Rev. D
1
3 x M20 or (optional)
/
in NPT
2
3
1 x M25 or (optional)
/
in NPT (remote system only)
4
1
/
NPT for
4
13 Transmitter Specification
1
1
/
in OD pipe (with
/
NPT option)
4
2
105