Endura AZ30 series integral probe and remote transmitter
Combustion oxygen monitor
HART Communications
Version
5.7 as standard
Integration
Device Type Manager (DTM) and Electronic Device Description (EDD)
Provide online / offline device configuration, online monitoring of measurement values and diagnostic states
DTM
FDT v1.2.1 compliant
Works with FDT framework packages (for example, ABB Asset Vision Basic)
EDD
Compliant with suitable framework tools (for example, SDC 625 and Simatic PDM tools)
Infrared Service Port
Accessibility
Through front face
Type
IrDA standard
Baud rate
Up to 115 k baud
Functions
Firmware update
Remote HMI
Diagnostic log download
Datalog output
HART via IrDA
Languages
English
French
German
Italian
Spanish
Calibration
Manual calibration
1 point (offset)
1 point (factor)
2 point (offset + factor)
Automatic calibration
1 point (offset)
2 point (offset + factor)
Calibration control
Front panel controls
Digital inputs
HART commands
User-defined schedule
Calibration scheduler
User-defined schedule enables automatic calibration frequency to be set from 1 day to 12 months
COI/AZ30E–EN Rev. D
13 Transmitter Specification
DS/AZ30-EN Rev. D
107