Cleaning And Disinfecting; Cleaning Ultrasound System Surfaces - Siemens Acuson S1000 Instructions For Use Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

Cleaning and Disinfecting

You must take all necessary precautions to eliminate the possibility of exposing patients,
operators, or third parties to hazardous or infectious materials. Use universal precautions when
cleaning and disinfecting. You should treat all portions of the ultrasound system that come in
contact with human blood or other body fluids as if they were known to be infectious.
All exterior parts of the ultrasound system, including the control panel, keyboard, transducers,
and biopsy devices, should be cleaned and/or disinfected as necessary or between uses. Clean
each component to remove any surface particles. Disinfect components to destroy pathogens
and other microorganisms.

Cleaning Ultrasound System Surfaces

WARNING: To avoid electric shock and damage to the ultrasound system, power off and unplug
the equipment from the AC power outlet before cleaning and disinfecting.
WARNING: Disinfectants and cleaning methods listed are recommended by Siemens for
compatibility with product materials, not for biological effectiveness. Refer to disinfectant label
instructions for guidance on disinfection efficacy and appropriate clinical uses.
WARNING: The use of any disinfectants other than those specified in the instructions for use
may damage the ultrasound system and accessory surfaces and, as a result, may create
electrical hazards for the patients and/or users.
Caution: To avoid the possibility of static shock and damage to the ultrasound system, avoid the
use of aerosol spray cleaners on the monitor screens.
Caution: Do not clean the ultrasound system with chlorinated or aromatic solvents, acidic or
basic solutions, isopropyl alcohol or strong cleaners such as ammoniated products, as these can
damage the surface of the system. Use the recommended cleaning procedure.
Caution: Do not use spray cleaners on the ultrasound system, as this may force cleaning fluid
into the system and damage electronic components. It is also possible for the solvent fumes to
build up and form flammable gases or damage internal components.
Caution: Do not pour any fluid onto the ultrasound system surfaces, as fluid seepage into the
electrical circuitry may cause excessive leakage current or system failure.
Caution: Prior to cleaning or disinfecting the system, you must disconnect and remove
transducers from the ultrasound system. Unintended contact with disinfectants approved only
for use with the system can result in damage to the transducers.
I n s t r u c t i o n s f o r U s e
2 S a f e t y a n d C a r e
2 - 2 9