Connecting System Accessories

The ultrasound system has connections for system accessories.

Footswitch

Attach the optional footswitch connector into the corresponding socket located on the front
system panel.

Physio Cables

Example of Physio panel.
Attach the optional Physio leads and auxiliary connectors to the Physio panel, located on the
left front of the system.
The optional Physio feature allows the system to display a scrolling ECG waveform on the
image screen.
Note: The Physio inputs are Defibrillation Proof. However, in the event of defibrillation while using
the Physio function, the Physio inputs may become saturated (overloaded). An ECG pattern may not
be visible for up to 30 seconds. After this time, the Physio function should return to normal operation.
To connect the Physio cables:
Connect the six-pin Physio cable to the socket labeled ECG on the front of the system.
I n s t r u c t i o n s f o r U s e
1
ECG connector
2
Aux 1 ECG plug
3 S y s t e m S e t u p
3 - 1 5