Figure 5-6 Analytical Module; Figure 5-7 Analytical Processor Board - ABB NGC8206 User Manual

Chromatograph
Hide thumbs Also See for NGC8206:
Table of Contents

Advertisement

11) Detach the analytical module rear face jack J1 and J4, if the auxiliary heater is
installed (see
12) Set module on a clean, lint-free surface.
13) Verify that the gasket on the feed-through assembly manifold interface is in
place, in good condition and free from metal filings or other contamination. If the
gasket has fallen off inside the enclosure or stuck to the GC module, replace
onto the feed-through manifold interface, ensuring that the gasket does not cover
the gas portholes.
14) Verify the S1 auxiliary heater switch is set to the correct position. If using the
auxiliary feed-through heater, set the position to Normal.
15) Insert the mounting screw into the analytical module.
Page 5–16
Figure
5-7).

Figure 5-6 Analytical Module

Figure 5-7 Analytical Processor Board

2101510 Rev. AG