Replacing Analytical Processor To Termination Panel Cable; Instructions - ABB NGC8206 User Manual

Chromatograph
Hide thumbs Also See for NGC8206:
Table of Contents

Advertisement

12) Reach into the enclosure through the front opening, and unplug the ribbon cable
from the rear of the termination panel J4.
13) On the replacement cable, verify the orientation by viewing the keyed receptacle
on the termination panel and cable. Insert the plug into the J4 connector.
14) Verify that the gasket on the feed-through assembly manifold interface is in place
and in good condition. If the gasket has fallen off inside the enclosure or is stuck
to the GC module, replace onto the feed-through manifold interface, ensuring
that the gasket does not cover the gas portholes.
15) Insert the mounting screw into the analytical module.
16) Holding the analytical module at the opening of the enclosure, reconnect jack J1
and J4, if the auxiliary heater is installed (see
17) Carefully insert the module into the enclosure, rotating the module to ensure that
the rear components clear the manifold interface on the inside area of the feed-
through assembly. The feed-through manifold interface and analytical module are
keyed to ensure the proper alignment.
18) When the analytical module is in place, tighten the mounting screw.
19) Use
20) Plug the termination panel into the digital controller ribbon cable and then into the
digital controller assembly.
21) Insert the lithium battery pack into the enclosure between the enclosure and the
thermal flask.
22) Once the unit is reassembled, apply power to the NGC (Step 6).
23) Reinstall the front and rear end caps.

5.23 Replacing Analytical Processor to Termination Panel Cable

Should the cable connecting the analytical processor to the termination panel
require replacement, use the following instructions.

5.23.1 Instructions

1) On the Analyzer Operation screen, click Hold under Next Mode. When the unit
completes the current cycle and enters hold, continue to the next step.
2) Collect data from the unit.
3) Back up the configuration files following the instructions detailed previously in the
section,
Page 5–30
Replacing Digital Controller Complete Assembly
Note that the termination panel to digital controller ribbon
cable pin 1 wire is not red. On the digital controller board, the
red edge (pin 1) of the cable should plug onto pin 50, the right
side of the plug. The plug is keyed; do not force plug into
connector.
Note that since the power was removed from this unit, the
NGC will perform start-up diagnostics and stabilize. If the user
has disabled the start-up diagnostics, they should be enabled
and power cycled to the unit. If the power has been withheld
from the unit for an unknown or lengthy period of time, a
complete start-up should be performed.
For more information on enabling the diagnostics in PCCU,
click Diagnostics and then Help.
Backing Up Configuration Files
Figure
5-6).
previously covered.
(Save).
2101510 Rev. AG