Enc82L Calibration Gas Bottle Installation; Materials; Instructions; Figure 3-40 Carrier Gas Low Pressure Switch Wiring Instruction - ABB NGC8206 User Manual

Chromatograph
Hide thumbs Also See for NGC8206:
Table of Contents

Advertisement

(+)
(+)
(-)
J2
2
1
INPUTS
DIGITAL I/0
NGC TERMINATION BOARD

Figure 3-40 Carrier Gas Low Pressure Switch Wiring Instruction

5) Using a small flat blade screw driver, loosen DI2 pins 3 and 4.
6) Insert the red wire into the (+) terminal (pin 3).
7) Retighten pin 3.
8) Insert the black wire into the (-) terminal (pin 4).
9) Retighten pin 4.
10) Replace the termination connector in the J2 board connector.

3.20 ENC82L Calibration Gas Bottle Installation

3.20.1 Materials

Strapping material (shipped with environmental enclosure)
Calibration gas blend bottle

3.20.2 Instructions

1) Locate the bottle bracket in the right front area of the environmental enclosure
(see
2) Set the calibration bottle inside of the enclosure, situated against the bottle
bracket.
3) Thread the strap through the holes in the bracket and around the bottle. Insert
the end of the strap into worm gear.
4) Using a flat blade screw driver, turn the screw on worm gear until strap is snug.
2101510 Rev. AG
(+)
(+)
(-)
(-)
(-)
1
2
OUTPUTS
The calibration gas bottle mounting rack is used to hold the
calibration gas bottle when located inside of the large
environmental enclosure. Refer to
A gas regulator should be installed on each gas bottle.
Figure
3-41).
BARRIER REQUIRED FOR ALL
INSTALLATIONS OTHER THAN
NON-HAZARDOUS
REGULATOR
ASSY
Figure
3-41.
CARRIER
CYLINDER
Page 3–39