Be certain the installation site is clean and free of foreign debris that could affect
NGC operation.
The NGC should be located as close as possible to the sample probe installation
point. This prevents the need for high gas flow rates through sample lines to
assure the analysis accuracy of the current sample.
If there is more than one stream being analyzed, locate the NGC in a central
location to all sample probe points.
3.1.4

Installation

The following information (see
procedures to follow depending upon the type of installation: pipe mount, free-
standing, wall shelf or one of the two environmental enclosures (ENC82) for
inclement climates. While a set of instructions maybe indicated for your
installation, it may not be required specifically.
2101510 Rev. AG
The NGC8200 is certified for installation in Classified
Hazardous locations. The heater and fittings in the
environmental enclosure may not have the same ratings. All
components of the installation, including accessories and
fittings must be approved for the classification rating of the
area of installation.

Table 3-1 Installation Matrix

Large Free-standing Environmental Enclosure
Small Free-standing Environmental Enclosure
Table
3-1) should help to determine the
Installation Instructions