GE CT-9800 Technical Publications page 427

Hide thumbs Also See for CT-9800:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
Rev. 3
4. Then we will Read Switch Register and
force a 'PANIC' by typing
I
(PANIC HANDLER ADDR.)R
The System should respond with a message indicating a 'CPU malfunction...'
P
UNIT 3 READY?
P
DONE
5. Set the tape drive thumbswitch back to 0.
6. Reset the CPU, Bring the disk drive to "READY". Bring the system to the 'R' level.
7. Enter
INIT MT0
8. Enter
DIR EDD;DUMP MT0:1 CTEX.SW CTSTAT
9. Enter
RELEASE MT0
10. Dismount the tape from the tape drive.
11. Please send the tape, a copy of the recorded information and any additional comments to :
GE Medical Systems
CT Service Engineering (NB918)
16800 W. Ryerson Rd.
New Berlin, WI 53151
11/
Type a capital I to reset the S140.
Type the panic handler address followed by a capital R.
Type a capital P to proceed
CPU message tells you to mount a tape on the drive and set the tape drive
select switch to unit #3 (thumbwheel switch).
Type a capital P to proceed. The CPU then writes an image of the core
memory to tape.
When CPU is done, remove the tape and set the tape drive select switch to
unit # 0.
8-4-7
------------------------------------
!
!
(PANIC HANDLER ADDR)
------------------------------------
CT 9800 QUICK SYSTEM
Direction 18000
!
!
<LF>