The INITIALIZER PROGRAM Continued:
The following example is for an HTD Dual Ace system only:
IMPORTANT: For HTD, you MUST select units 0, 1 and 3 and perform [Z]DKINIT on each.
Otherwise, the RDOS and UNIX partitions will not be properly initialized.
However, Unit 2 does not have to be selected if it is ready.
GESCSI Disk Initializer
Copyright 1987, Zetaco Inc.
[E]nter a device
[S]tart a device
[R]estart the program
[C]ommand list
[F]lags
[X]eq and error log
Enter a Command Selection (C=Cmd List): E<CR>
Enter the device code [24]: <CR>
Start initialization of controller.
Micro-code rev #: 3.00
End initialization of controller.
Unit
0. is Ready; Select (YES,[NO]): YES<CR>
Command choices:
[Z]DKINIT ( ZDKINIT, INSTALL MICROCODE, INIT SOFT LOG)
[A]NALYZE DISK
[R]ELOCATE BLOCKS
[M]ICRO CODE INSTALL ONLY
ENTER YOUR SELECTION [Z]: <CR>
UNIT 0. IS SELECTED
UNIT 1. IS READY Select (YES,[NO]): YES<CR>
Command choices:
[Z]DKINIT ( ZDKINIT, INSTALL MICROCODE, INIT SOFT LOG)
[A]NALYZE DISK
[R]ELOCATE BLOCKS
[M]ICROCODE INSTALL ONLY
ENTER YOUR SELECTION [Z]: <CR>
UNIT 0. IS SELECTED
UNIT 1. IS SELECTED
UNIT 2. IS NOT READY
UNIT 3. IS READY Select (YES,[NO]): YES<CR> Note: Units 0, 1, and 3 MUST be selected for
HTD.
Command choices:
[Z]DKINIT ( ZDKINIT, INSTALL MICROCODE, INIT SOFT LOG)
[A]NALYZE DISK
[R]ELOCATE BLOCKS
[M]ICROCODE INSTALL ONLY
ENTER YOUR SELECTION [Z]: <CR>
**** Note **** You must enter [S]tart in order to have the selected operation
actually take place.
Enter a Command Selection (C=Cmd List): S<CR>
Start all entered devices ([YES] NO): <CR>
Rev. 4.00
Command List
____________
[D]elete a device
[H]alt a device
[L]ist device status
[P]rinter control
[M]ode of display
[Q]uit
Note: Use [Z] without selecting any
units to initialize the controller only.