GE CT-9800 Technical Publications page 452

Hide thumbs Also See for CT-9800:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
Rev. 3
*** Deadstart Program (DS) *** Revision 1.05
Configuration Information
Processor Type = ST-X
01 Compute Head(s)
02 Back Projector Board(s)
ACU Rounding Mode = ST-100
SMP MUX NOT PRESENT
Main Memory 1.0 Mega words
*** RESET Complete ***
*** Deadstarts Complete *** Unable to establish communications with AP.*E*
*** Deadstart Program (DS) *** Revision 1.05
Configuration Information
Processor Type = ST-X
01 Compute Head(s)
02 Back Projector Board(s)
ACU Rounding Mode = ST-100
SMP MUX NOT PRESENT
Main Memory 1.0 Mega words
***
RESET Complete
PASSED Master ROM test
PASSED Master private RAM test
PASSED Master public memory test
PASSED Master TIMER test
***
FAILED Slave CP test
***
FAILED Slave ROM test
*E*
***
FAILED Slave private RAM test, failing address = 0000
*E*
***
FAILED Slave public memory test, failing address = 0000
***
RESET Complete ***
Deadstart Part 2 Download in Progress
........
PASSED IOP0 test
*** FAILED PLS Head
*** FAILED Backprojector Controller test
failing Back Projectors
00
01
02
Main memory = 1.0 Mega words found
IOP's are = DAS CHI - - -
*** Deadstarts Complete ***
***
0 test
03
04
05
*** NO IOC BOARD ***
*** NO SLAVE BOARD ***
8-6-14
CT 9800 QUICK SYSTEM
Direction 18000

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents