2 - 2
7.
Make sure the capacitor bank is discharged.
WARNUNG
WARNING
8.
Remove the lower and upper arm covers (pos. 4 / Fig. 1).
NOTICE
MOBILETT Plus/ -E/ -HP
The capacitor bank might still contain up to 350 VDC.
Life-threatening hazard of electric shock.
Follow the safety procedures described in the docu-
ment Service Instructions for detailed instructions on
how to ensure the capacitor bank is discharged.
For MOBILETTs with remote exposure switch system
installed:
When removing the lower arm cover, be sure to unplug
the module connector for the remote control from the
electronics package alongside the main MOBILETT
CPU board (see pos. 1 / Fig. 18). Dismount the IR sen-
sor from the lower arm cover.
Remember how the cable was attached for future steps
in section "Cable assembling"!
SPR8-220.031.04
Rev. 04
Page 2 of 30
09.03
CS PS 24
Installation
Siemens AG
Medical Solutions