Installation Of Power Supply, Mobilett Plus Hp Only - Siemens MOBILETT Plus Installation Instructions Manual

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

2 - 18

Installation of power supply, MOBILETT Plus HP only!

1.
For MOBILETT plus HP with a tilt bar, remove the tilt bar (see Fig. 19).
Fig. 19 Removal of tilt bar
2.
Remove pin plugs K201, K202 and K203 on board D201.
3.
Remove the front screw (pos. 1 / Fig. 20).
4.
Remove the two back screws (pos. 1 / Fig. 21).
Fig. 21 Board D201, back
5.
Remove board D201 including the aluminum plate and place it on a work bench.
CAUTION
MOBILETT Plus/ -E/ -HP
MOB00644
1
Sensitive electronic equipment!
Improper handling could cause permanent damage to
the circuits.
Follow ESD guidelines while working with the alumi-
num plate and mounting the new power supply and be
careful not to damage board D201.
SPR8-220.031.04
Rev. 04
1
Fig. 20 Board D201, front
Page 18 of 30
09.03
CS PS 24
Installation
2
Siemens AG
Medical Solutions

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus eMobilett plus hp