2 - 28
14.
Place the cable as shown in Fig. 39. The cable should run underneath the flat
conductor cables attached to board D1.
15.
Fasten the cable with a cable tie (2.5x100mm, pos. 1 / Fig. 39).
16.
Pin plug K23 is threaded through the hole (pos. 2 / Fig. 40).
17.
Fasten the cable with a cable tie (2.5x100mm, pos. 1 / Fig. 40). Make sure there is
no slack.
18.
Connect pin plug K23 to pin plug K23 of the cable of the new power supply.
19.
Continue installation under section "Final assembly".
MOBILETT Plus/ -E/ -HP
1
Fig. 39 Placing cable
1
2
Fig. 40 Threading cable
SPR8-220.031.04
Rev. 04
K9
D1
Page 28 of 30
09.03
CS PS 24
Installation
Siemens AG