Reassembly - Siemens MOBILETT Plus Installation Instructions Manual

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

Installation
9.
Place the cable according to Fig. 16. Make sure the cable runs underneath the flat
conductor cables connected to board D1.
(to K21, K22)
(to K23)
Fig. 16 Placing cable
10.
Connect K9a to K9 on board D1 (Fig. 16).
11.
Remount the protection plate over board D1.

Reassembly

1.
Remount the collimator covers "A" and "B" (see Fig. 2).
2.
Turn in the stop screws (Part No. 60 32 965) half way in the collimator.
3.
Place the shadowcross in position with the aid of the two stop screws and mount
the new collimator cover "C" (see Fig. 2).
4.
After tightening two of the screws that hold the collimator cover "C", the stop
screws can be replaced with the remaining original screws.
NOTICE
5.
Mount the adjustment knobs.
6.
Connect the pin plug K22 to K22 in the new upper arm cover (pos. 1 / Fig. 7).
7.
Mount the new upper arm cover.
8.
Mount the left and right circular covers on the upper and lower joint (pos. 5 / Fig.
1).
Siemens AG
Medical Solutions
For MOBILETTs with field size indicator installed:
Use the label kit (Part No. 65 28 108) and attach the
labels according to the old cover.
Close the collimator lids and mount the adjustment
knobs so that the "0"s on the knob scales are facing
the arrows.
Check the mounting by opening the collimator lids and
close them again completely. The "0" on the knob scale
should be just in front of the arrows.
SPR8-220.031.04
Rev. 04
K9a
D1
Page 15 of 30
09.03
CS PS 24
2 - 15
K9b
MOBILETT Plus/ -E/ -HP