6.
If needed, drill three holes with a diameter of 5 mm in the aluminum plate as
shown in Fig. 22 for the new power supply unit. The distances are: A = 65 mm, B =
46 mm, C = 32.5 mm, D = 108 mm and E = 25 mm. Also drill a hole (5 mm
diameter) as shown in pos. 1 / Fig. 22 to attach the cable to the new power supply.
CAUTION
Battery charger
7.
Cut the cable ties as shown in Fig. 23.
Siemens AG
Medical Solutions
If drilling is necessary:
Drilling chips can cause injuries to eyes.
Use proper protection.
Fig. 22 Adjustments for new power supply
Fig. 23 Cutting cable ties
SPR8-220.031.04
Rev. 04
E
D
C
B
A
1