Installation Of Power Supply, Mobilett Plus/ -E Only - Siemens MOBILETT Plus Installation Instructions Manual

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

2 - 24

Installation of power supply, MOBILETT Plus/ -E only!

1.
For MOBILETT Plus/ -E with a tilt bar (see Fig. 31), remove the tilt bar.
2.
If needed, drill two holes of 4.5 mm diameter as shown in Fig. 32. The distances
are: A = 65 mm, B = 40 mm and C = 10 mm. This is only necessary for some older
units.
CAUTION
3.
Unplug pin plug K52 from board D8 (pos. 1 / Fig. 33).
4.
Fasten the cable shielding connector with the screw holding the protective cover
over board D8 (pos. 4 / Fig. 33). Replace the previously mounted washer with a
washer (KB 5.1mm) included in the kit.
5.
Attach pin plug K52b to pin plug K52. Then attach pin plug K52a to K52 on the
board (pos. 1 / Fig. 33).
MOBILETT Plus/ -E/ -HP
Fig. 31 Removing tilt bar
A
C
B
Fig. 32 Adjustments for new power supply
If drilling is necessary:
Drilling chips can cause injuries to eyes.
Use proper protection.
SPR8-220.031.04
Rev. 04
MOB00644
Page 24 of 30
09.03
CS PS 24
Installation
2
Siemens AG
Medical Solutions

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mobilett plus eMobilett plus hp

Table of Contents