Installation

Removal of covers

1.
Remove both the right and left circular covers (pos. 5 / Fig. 1). Unplug K1 from K1
on D1 (pos. 2 / Fig. 18) and remove the cover of the articulated shaft (pos. 6 / Fig.
1).
Fig. 2 Removal of collimator covers
2.
Turn the single tank so that the collimator is facing upwards (Fig. 2).
3.
Secure the arm and the single tank with the belts (Fig. 2).
4.
Remove covers "A" and "B" (Fig. 2).
5.
Remove the adjustment knobs according to Fig. 3.
1.
Remove collimator cover "C" (Fig. 2) and the shadowcross inside.
Siemens AG
Medical Solutions
C
A
B
Fig. 3 Removal of collimator cover "C"
SPR8-220.031.04
Rev. 04
1
2
Page 3 of 30
09.03
CS PS 24
2 - 3
2
MOBILETT Plus/ -E/ -HP