5.5
ISE A/D Conversion
(1)
ISE A/D Conversion Timing
Na
K
1
(a)
As shown in the above figure, measure the electromotive force 5 times each for Na,
K and Cl in this order.
(b)
The A/D conversion interval time for each is 10 msec.
(c)
ADC alarm is issued if the ADC count, even once, is 0.
(2)
ADC Setting Parameters
•
Number of integrations :
•
Software trigger interval :
(3)
ISE ADC Calibration Timing
2 V
8 V
(a)
As shown in the above figure, measure the reference voltages
(2 V, 8 V and offset voltage).
(b)
The A/D conversion interval time for each is 10 msec.
(c)
At the time of initialization and T/M resetting, the RESET command is issued for the
ADC controller.
(d)
Number of integrations (1 time) = 8
10 msec
Cl
REF
Na
10 msec
OFF
K ................................ REF
2
8
10 msec
5 - 8
Na
K
Cl
REF
5