Automatic analyzer
Table of Contents

Advertisement

Designation
No.
of Check
2
Assay code
(2) Photometric points after
check
3
Calibration
The following are checked
parameter
for measurable (level 1)
check
photometry tests.
(1) Relationship between
(2) Entry must be made for
(3) Entry must be made for
Details of Check
the specified reaction
time must not be set.
Refer to Note 6 in
"analytical method
table" of 1.1.1.
assay code and
calibration type
Refer to "relationship
between calibration type
and analytical method"
in 2.1.
CALIB. POINTS
necessary for calibration
type.
Refer to "output by each
calibration method and
check table" in 2.7.
STD POS. necessary
for calibration.
For STD specified by
measured STD No. in
"output by each
calibration method and
check table" of 2.7, it is
checked whether STD
POS. is input or not.
Alarm
Display
Subdivision
CHEM
1 to 37
PARAM?
(CH No.)
CLB. PARAM?
1 to 37
(CH No.)
2 - 31
(cont
d)
'
Remarks
NOTES:
1. Checked
against the
specification of
calibration test
selecting
information.