R.probe Rotation Adjustment; Stirring Rod Rotation Adjustment - Hitachi 902 Service Manual

Automatic analyzer
Table of Contents

Advertisement

(5)

R.probe Rotation Adjustment

(a)
Details of operation
Probe is not moved vertically.
(b)
Extent of operation
1) Rotation angle: Reference value (pulse count) is subjected to no. of correction
(6)

Stirring Rod Rotation Adjustment

(a)
Details of operation
Stirring mechanism is not moved vertically.
(b)
Extent of operation
1) Rotation angle: Same as in operation.
Above cell
Repeat
Rinse bath
pulses set on System Parameters screen, and corrected pulse
count shall be used.
Above cell
Repeat
Above reagent bottle (outer row)
Above reagent bottle (inner row)
Conditions for stopping
1) When STOP key is pressed.
2) When mechanism STOP level alarm occurs.
In case of 1), probe returns to rinse bath and stops.
Rinse bath
Conditions for stopping:
1) When STOP key is pressed.
2) When mechanism STOP level alarm occurs.
In case of 1), rod returns to rinse bath and stops.