Alarm
Alarm
Control
No.
5271 to
SERUM
INDEXES?
5356
5431
ON BOARD?
5441
REAG.POS?
5481
CLB(IS)
POS.
5511 to
ACI ERROR
5516
Sub-
Category
Level
code
104
1 to 36
WARNING (1) Although the sub-code
(ch.)
106
1
WARNING There is no measurable
107
1 to 38 WARNING (1) The reagent position
108
1
WARNING When ISE is provided,
114
1 to 6 WARNING Barcode IC malfunctions.
2 - 19
Description
corresponds to the
serum index
measurement test, the
rate-A assay is not
assigned.
(2) Although the sub-code
corresponds to the
serum index
measurement test and
sample blank is to be
corrected, reagent 2
discharge is specified.
(Analysis does not start.)
channel.
There is no channel for
which necessary reagent
has been prepared.
(Analysis does not start.)
specified for a
photometry assay is
also specified for other
photometry assay.
(2) The same reagent
position is specified for
both carry-over cleaning
agent and photometry
assay or ISE test.
(Analysis does not
start.)
calibrator or control
positions are set at 55 to 57.
(Analysis does not start.)
1: Unassigned
2: Unassigned
3: Sample
4: Unassigned
5: Unassigned
6: Transfer
(cont'd)
Remedy
Check parameter
for serum indexes.
Check each
parameter.
Check each
reagent.
Check the position
for control
calibration.
Replace RSDIST
PC board.