Hitachi 902 Service Manual page 31

Automatic analyzer
Table of Contents

Advertisement

Alarm
Alarm
Control
No.
5271 to
SERUM
INDEXES?
5356
5431
ON BOARD?
5441
REAG.POS?
5481
CLB(IS)
POS.
5511 to
ACI ERROR
5516
Sub-
Category
Level
code
104
1 to 36
WARNING (1) Although the sub-code
(ch.)
106
1
WARNING There is no measurable
107
1 to 38 WARNING (1) The reagent position
108
1
WARNING When ISE is provided,
114
1 to 6 WARNING Barcode IC malfunctions.
2 - 19
Description
corresponds to the
serum index
measurement test, the
rate-A assay is not
assigned.
(2) Although the sub-code
corresponds to the
serum index
measurement test and
sample blank is to be
corrected, reagent 2
discharge is specified.
(Analysis does not start.)
channel.
There is no channel for
which necessary reagent
has been prepared.
(Analysis does not start.)
specified for a
photometry assay is
also specified for other
photometry assay.
(2) The same reagent
position is specified for
both carry-over cleaning
agent and photometry
assay or ISE test.
(Analysis does not
start.)
calibrator or control
positions are set at 55 to 57.
(Analysis does not start.)
1: Unassigned
2: Unassigned
3: Sample
4: Unassigned
5: Unassigned
6: Transfer
(cont'd)
Remedy
Check parameter
for serum indexes.
Check each
parameter.
Check each
reagent.
Check the position
for control
calibration.
Replace RSDIST
PC board.