Designation
No.
of Check
5
Test-to-test
For formula No. where
compensa-
compensated test is
tion check
measurable:
(1) The compensation test
(2) The compensated test
(3) When the compensated
6
Serum index
(1) The assay code for tests
analytical
method
check
(2) In the RATE-A test with
7
Cell carry
(1) More than 2 types of
over
(2) Sub-codes (1 to 10)
8
ISE carry
R1 type is not specified for
over
ISE test of Reagent Probe
Carry Over Evasion.
Details of Check
must be measurable.
In addition, when the
compensation test is
already registered as a
compensated test in any
preceding formula No.,
the check result for the
formula No. must be
OK.
must be on the right
side of the formula.
test is a photometry
test, no ISE test must
be registered for the
compensation test.
for serum index
measurement must be
RATE-A. This check is
made only when the
tests are measurable.
serum indexes, no R2
must be specified when
sample blank correction
is carried out.
carry over evasion (cell)
are specified for 1 test.
signify the evasion
types.
Alarm
Display
Subdivision
CMP. TEST?
44 to 51
(FORMULA
No.)
S. INDEXES?
1 to 37
(CH No.)
CELL. C. O.
1 to 10
ISE C. O.
1 to 40
2 - 33
(cont
d)
'
Remarks
This check is not
performed in case
of ORIGINAL ABS.
Check and correct
the carry over
evasion (cell).
Check and correct
the Reagent Probe
carry over evasion.