7.2 Measurement

(1)
Forty-nine cells are measured in each cycle.
(2)
Thirty-nine measurements (4 cell blank measurements, 35 sample absorbance
measurements) are made in each cycle.
(3)
In case of at least one ADC failure in thirty-nine measurements, data is 0 and ADC alarm
is added.
(4)
ADC
(a)
Time: Refer to the description of ADC parameters.
(b)
Calibration equation
ABS in terms of 10 mm cell
(measured with 6 mm cell)
A = Calibrated ADC count for 2 V voltage
B = Calibrated ADC count for 6 V voltage
C = Measured ADC count
(c)
Immediately after calibration, subtraction is made atB initialization, resetting upon
operation start and T/M resetting. Note, however, that data at initialization is
discarded.
At this time, ADC CALIB alarm is issued if A or B is out of the range given below.
7546 ≤ A ≤ 8341
22640 ≥ B ≥ 25023
(d)
Number of conversion bits: 16 bits
(e)
The RESET command is issued to the ADC controller at initialization and T/M
resetting.
16667
×
=
100 000
,
X
B A
7 - 13
63338
B
25000
A
B A