Memory Initialization; Memory Switch Setting 1 - Hitachi 902 Service Manual

Automatic analyzer
Table of Contents

Advertisement

(6)

Memory Initialization

[Memory Initialization Menu]
[Menu End]
[Image Data Initialization]
Select [Image Data Initialization].
Is it desired to initialize image data
memory?
[Yes]
[No]
Select [Yes].
The image data memory is initialized.
"Initialization is under execution." is displayed during initialization.
Upon completion of initializing the image data memory, execute installation according to
"Procedure for Installation of Product Program".
(7)

Memory Switch Setting 1

(NOTE: Each switch is set as shown below. Setting change may disable startup of the
instrument. So avoid changing the setting.)
[Keying Sound]
[Buzzer Sound]
[Back Light Off Time] 10 Min Screen
[Automatic Commu-
nication Recovery]
Press [Write Check] key.
(8)
Memory Switch Setting 2
(NOTE: Each switch is set as shown below. Setting change may disable startup of the
instrument. So avoid changing the setting.)
[Data Bit Length] 8 Bits
[Stop Bit Length] 2 Bits
[Parity]
[Flow Control]
[Baud Rate]
[Lamp Bit Control] Invalid
Press [Write Check] key.
OFF
Stop
ON
Next
Write
No
Check
Stop
Previous
Even
Screen
RS/CS
9600 bps Write
Check
12 - 20