Specifications Of Deionized Water Supply Unit For Model 902 - Hitachi 902 Service Manual

Automatic analyzer
Table of Contents

Advertisement

NOTE: Specifications of Deionized Water Supply Unit for Model 902
Described below are the specifications for connection of the deionized water supply
unit to the Model 902.
Water quality
Water supply capacity : 20 L/hr
Water pressure :
Main unit intake
Control signal
Deionized
water supply
unit
Wiring/piping
: Neutral water containing no substances that adversely affect
measurement, such as deionized water and distilled water
50-340 Kpa; 0.5-3.5 kgf/cm
germ-free
: Connectable to nipple of 16 mm in diameter and 28 mm in
length
: Water aspiration via 100 V AC contact signal
1
5
Connector (Hirose RM15QPH-8-P)
: Within 5 m long
12 - 5
; Conductivity < 1 µs/cm or less
2
100 V AC
902 main unit