2.4.4 ISE Data Alarms

Alarm on ISE data processing has 2 kinds shown below; calibration alarm and data alarm.

(1) Calibration Alarm

Priority
Alarm
1
STD error
2
Slope
abnormal
3
Electrode
preparation
4
Internal
standard
concentration
abnormal
5
Calibration
error
< Key >
N
:
Routine sample measurement
E
:
Stat sample measurement
C
:
Control sample measurement
STD :
Calibration measurement
Printout
CRT
S. I/F
STD?
S
SLOPE?
E
PREP
R
I. STD
D
CALIB
B
ISE
Processing
N.E.C.R
STD
×
O
×
O
×
O
×
O
×
O
2 - 49
Alarm Output
Remarks
on Operation
Monitor Screen
O
In calibration
measurement,
this alarm is set
at any of the
data alarms in
(2) (excluding
data alarms 6 to
9).
O
O
O
O