Code
Error Message
1430
The image cannot be saved
directly from a shared network
folder.
Assigning a drive letter to the
connection is required to save
the image using My Computer.
1450
Failed to save the image.
1470
Added items exceed the
maximum number of the items
that can be added on an
image.
No more items can be added.
1471
Proceeding with this operation
will delete all the items.
Do you want to delete?
1510
Cannot evacuate.
Check if the stage is closed
right.
Check if the diaphragm pump
is working.
Start up the main unit and
application again.
Contact service if the failure
occurs repeatedly.
Countermeasure
This message appears if you attempt to save an
image directly from My Network. Click the [OK]
button to display the data saving window again.
No image can be saved directly from My
Network. For image saving, assign a network
drive on My Computer and specify the network
drive on the data saving window.
Image saving failed because any error occurred
during image saving. Click the [OK] button to
close the message. Make sure the saving
destination allows your access and it is not
write-protected (not a read-only memory). If the
same error occurs repeatedly, contact the
nearest service representative.
This message appears when the number of data
items exceeds 50. Click the [OK] button to close
the message.
To add new items, delete unnecessary items
first.
This message appears to confirm the operation.
Click the [OK] button to delete all items. Or, click
the [Cancel] button to cancel the operation.
A time-out (120 s) has occurred during
evacuation.
Click the [OK] button to close the message.
Restart the system by the following procedure.
1. Turn off the PC power.
2. Turn off the main unit power.
3. Restart the PC.
4. Turn on the main unit power and start the
application.
Make sure the specimen stage and the electron
gun are securely closed. Turn on power for the
main unit while pressing the electron gun
against the specimen stage. Also, make sure the
vacuum tube for the diaphragm pump is not bent
or kinked.