Advertisement

4. Start of 3D-Image Capture
With the PC control application program started, select 3D-Image Capture from the File
menu and open the 3D-Image Capture dialog.
NOTE : When the SEM main unit is under the following states, 3D-Image Capture cannot be
executed.
 Accelerating voltage is OFF
NOTE : When [SE] or [Mix] is selected from [Signal Select], it is not possible to select
[3D-Image Capture].
Select the image saving mode in the 3D-Image Capture dialog, and execute capturing.
 [Quick Save]
Image size: 640  480 (Image only)
Image acquisition time: The capture of about ten seconds per one image is executed four
 [Save]
Image size: 1280  960 (Image only)
Image acquisition time: The capture of about 40 seconds per one image is executed four
Figure 7.14.1-3 3D-Image Capture of File Menu
Figure 7.14.1-4 3D-Image Capture Dialog
times. It takes about 40 seconds in total.
times. It takes about 160 seconds in total.