DIGitize
DIGitize
command
The
DIGITIZE
command
is
used to acquire waveform data for transfer
over
the
HP-IB. It
causes
an acquisition to take place on the specified
channel(s) with the resulting data being placed in the channel buffer.
The
ACQUIRE
subsystem commands are used to
set
up conditions
such
as
TYPE,
number of
POINTS,
and the
COUNT
for the
next
DIGITIZE
command. See the
ACQUIRE
subsystem for
a
description
of
these
commands.
To
determine the actual number of points that
are acquired
and how the data will be transferred,
refer
to the
WAVEFORM
Subsystem commands.
For
more information on
the
DIGITIZE
command refer
to
the section
on the
DIGITIZE
command
in the chapter
"Introduction to
Programming an
Oscilloscope."
Note
Sending
the
DIGITIZE
command will
turn
off
any
unused
channels.
When the Digitize operation
is
complete the instrument
is
placed in
the
stopped mode.
When
the instrument
is
restarted, with
a
RUN
command
or the front panel
RUN
key,
the digitized data
stored in the channel
buffers will be overwritten.
Therefore,
ensure all operations that
require
the
digitized data are completed before restarting the
HP
54501A.
The
speed
of
the total digitize operations may be improved
if
two or more
DIGITIZE
commands are
sent
without changing other parameters.
The
sources
for
the
:DIGITIZE
command
are channels
1
through 4.
Command
Syntax:
:DIGitizc CHANnel
<
N
>
[,CHANnel
<
N
>
]...
Where:
<N>
::=
1,2,
3,
or
4.
Example:
OUTPUT 707;":DlGmZE CHANNELI.CHANNELZ"
HP 54501A
Programming
Root Level Commands