Vstart - HP 54501A Programming Reference Manual

Digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

VSTArt
VSTArt
command/query
The
:MEASURE: VSTART
command moves
VMarker
1
to the specified
voltage.
The
values
are limited to the currently defined channel, function,
or memory range.
The
VSTART
query returns the current voltage
level
of
VMarker
1.
Note
The
short
form
of
this
command
does
not follow
the
defined
convention.
The
short
form
"VST
is
the same
for
VSTART
and
VSTOP,
so be
careful not to send
this
form for
the
VSTART
command. Sending "VST" will
produce
an
error.
Command
Syntax:
:MEASure:VSTArt
<
voltage
>
Example:
OUTPUT
707;":MEAS:VSTA -10MV"
Where:
<
voltage
>
::
=
voltage
value for
Vmarker
1
Query Syntax:
:MEASure:VSTArt?
Returned Format:
[:MEASure:VSTArt]
<
value
>
<NL>
Where:
<value>
::
=
voltage at VMarker
1
(exponential
-
NR3 format)
Example:
DIM vst$[50]
OUTPUT
707;":MEASURE:VSTART?"
ENTER 707;
Vst$
PRINT Vst$
HP
54501A
Programming
Measure Subsystem
14-61