TSTOp
TSTOp
command/query
The
:MEASURE:TSTOP
command moves the stop
marker
to the
specified time with
respect to the
trigger time.
The TSTOP query
returns the time
at the stop
marker.
Note
The
short
form
of
this
command
does
not follow the
defined
convention. The short
form
"TST
is
the same
for
TSTART
and
TSTOP,
so be
careful not to send
this
form for
the
TSTOP
command. Sending "TST' will
produce
an
error.
Command
Syntax:
:MEASure:TSTOp
<
stop
marker time
>
Where:
<
stop marker
time
>
::
=
time
at stop marker
in
seconds
Example:
OUTPUT 707;":MEAS:TSTOP
40
NS"
Query Syntax:
:MEASure:TSTOp?
Returned Format:
[:MEASure:TSTOp]
<vaiue>
<NL>
Where:
<
value
>
::
=
time at
stop
marker
in
seconds (exponential
-
NR3
format)
Example:
DIM Tst$[50]
OUTPUT 707;":MEASURE:TSTOP?"
ENTER
707;Tst$
PRINT
Tst$