Define - HP 54501A Programming Reference Manual

Digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

DEFine
DEFine
command/query
The
:MEASURE:DEFINE
command
sets
up the definition
for
a
measurement.
The
DEFINE
query returns the current setup.
Command
Syntax:
:MEASure:DEFine
<
measurement_spec
>
Where:
<
measurement_spec
>
::=
{DELay
<polarity>, <edge_num>, <level>,
<
polarity
>,
<edge_num>,
<
level
>
|
PWIDth
{MIDDIe
|
UPPer
|
LOWer}
|
NWIDth
{MIDDIe
|
UPPer
|
LOWer}}
Where:
<polarity> ::= {POSitive |
NEGative}
<edge_num
>
::
=
an integer, -127 to 127 (excluding
0)
specifying
a
displayed edge
<
level
>
::= {MIDDIe |
UPPer
|
LOWer}
Example:
OUTPUT 707;":MEAS:DEFINE DELAY,POS1TIVE,1,UPPER,NEGATIVE,2,MIDDLE"
This example will
set
the parameters
for
a
time measurement from the
first
positive edge
at
the upper threshold level to the second negative edge
at
the middle threshold.
If
one source
is
specifed, both parameters apply
to
that signal.
If
two sources are specified the measurement
is
from the
first
positive edge on
source
1
to the
second negative edge on source 2.
HP 54501A
Programming
Measure Subsystem
14-15

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents